QuantiFluor® dsDNA System for High Concentration DNA
Promega Corporation
Publication Date: June 2013 tpub123
Abstract
The QuantiFluor® dsDNA System is a highly accurate fluorescent method for quantifying DNA. While the technical manual indicates a dye dilution of 1:200 to obtain linear standard curves for up to 200ng of DNA per assay, the chemistry can be used to quantitate significantly higher amounts of DNA by altering the dilution factor of the QuantiFluor® dsDNA dye. A QuantiFluor® dsDNA dye dilution of 1:25 resulted in highly linear data for up to 2.5µg of DNA in a 200µl final reaction volume. This enables quantitation suitable for many current extraction methods without the need for sample dilution prior to the assay.
Introduction
The QuantiFluor® dsDNA System contains a fluorescent DNA-binding dye that is used to quantitate double-stranded DNA (dsDNA) without binding to contaminating single stranded DNA (ssDNA), RNA, or protein. This highly specific method is useful for quantifying small amounts of DNA with a detection limit of 50pg/ml, greatly out performing absorbance measurements at 260nm. Because of the simple assay design, the QuantiFluor® dsDNA System is easily formatted for 96-well plates or other higher throughput formats.
When using the high-concentration dsDNA curve as specified in the technical manual (TM346), results are linear between 100pg and 200ng of DNA per 200µl assay. The technical manual indicates a dye dilution of 1:200 for use with this range of DNA. In this article, we show that by changing the dilution of QuantiFluor® dsDNA Dye to 1:25, the linear range of dsDNA detection can be increased to approximately 2.5µg DNA/assay without losing linearity in lower ranges of dsDNA while maintaining reliable quantitation of DNA at much higher concentrations than previously possible. For example laboratories isolating DNA from large sample volumes where a sample input of precisely 2µl may be required, the use of a 1:25 dye dilution enables quantitation of up to 1.25µg/µl undiluted sample DNA (2.5µg DNA in assay).
Materials
- QuantiFluor™ dsDNA System (Cat.# E2670)
- Human Genomic DNA (Cat.# G3041)
- GloMax®-Multi+ Microplate Reader (Cat.# E9032)
Methods
The general protocol TM346 was modified to incorporate varied DNA concentrations, dye dilutions, and standard type as specified below.
In order to determine the linear range for various dye dilutions of QuantiFluor™ dsDNA dye, Human Genomic DNA was prepared in 1X TE Buffer for 0.25µg to 3.0µg in a final volume of 100µl. QuantiFluor® dsDNA dye was diluted 1:200, 1:100, 1:50, 1:30, 1:25, 1:20, 1:15, and 1:10 in 1X TE Buffer to a final volume of 100µl and mixed with DNA solutions for a final reaction volume of 200µl. Reactions were mixed on a plate shaker for 1 minute, covered from light and incubated at room temperature for 5 minutes and then analyzed with the Blue Fluorescence Optical Kit (490nmEx/510–570nmEm) on the GloMax®-Multi+ Microplate Reader in a 96-well plate. All reactions were performed in quadruplicate. Data were analyzed with Microsoft Excel to determine linearity, with r2 values above 0.99 considered to be highly linear.
Results
Dye dilutions of 1:200 and 1:100 resulted in a plateau of RFU values for greater than 250ng and 500ng of DNA, respectively (Figure 1). Dilutions of 1:50 and 1:30 improved linearity slightly but showed increasing variability for DNA amounts higher than 1.25µg and 2.0µg, respectively (Figure 1).
Dye dilutions of 1:25 and 1:20 resulted in the broadest linear ranges of all dye dilutions tested. The 1:25 dilution was linear up to 2.5µg, and the 1:20 dilution was linear for up to 2.0µg (Figure 2). Dye dilutions of 1:15 and 1:10 did not improve upon this range and resulted in increasingly variable values for higher than 2.0µg DNA.
We calculated r2 values for all dye dilutions (Table 1), and values greater than 0.99 were considered to be linear. The 1:25 dilution showed the greatest range and detected up to 2.5µg of DNA per assay without loss of linearity.
Conclusion
The QuantiFluor® dsDNA System was used to quantitate high concentrations of DNA by altering the dilution of QuantiFluor® dsDNA dye used. A dilution of 1:25 enabled accurate quantitation of up to 2.5µg of DNA per assay with linear results. Further concentrating the dye did not improve upon the linear range at concentrations beyond 2.0µg of DNA, and an increase in standard deviation and background fluorescence was observed for these samples.
Related Products
Related Protocols
Frequently Asked Questions
How to Cite This Article
Scientific Style and Format, 7th edition, 2006
Truman, A and Wieczorek, D. QuantiFluor® dsDNA System for High Concentration DNA. [Internet] June 2013 tpub123. [cited: year, month, date]. Available from: https://www.promega.com/es-es/resources/pubhub/quantifluor-dsdna-system-for-high-concentration-dna/
American Medical Association, Manual of Style, 10th edition, 2007
Truman, A and Wieczorek, D. QuantiFluor® dsDNA System for High Concentration DNA. Promega Corporation Web site. https://www.promega.com/es-es/resources/pubhub/quantifluor-dsdna-system-for-high-concentration-dna/ Updated June 2013 tpub123. Accessed Month Day, Year.