To be certain that your Spectrum Compact CE System is ready to go from the start, Installation Qualification (IQ) provides a series of formal on-site instrument checks, delivers written documentation about the functionality of your Spectrum Compact CE System and demonstrates that everything ordered with your instrument is supplied and installed in your laboratory. Upon delivery to your lab, the instrument and its components will be visually inspected and reviewed for completeness. Following the inspection, the instrument will be powered on to confirm that the system is properly functional.
- Installation by qualified Promega personnel
- Inspection of shipping containers, instrument and accessories
- Comparison of items received versus items on your Purchase Order
- Inspection of laboratory conditions
- Review of all hazards and precautions with users
- Confirmation or installation of correct hardware and firmware version
- Diagnostic self-test of your instrument
- Record of installation and actions
- Service includes site visit, instrument sticker and IQ documentation